MIELE – voltage source with 50/60 Hz for the simulation of worldwide mains voltages

Schuntermann has developed, produced and commissioned two basic supply units and 21 laboratory supply units after coordinating the task with Miele.

The outputs of the control cabinet distributor are three-phase voltage regulators with autowinding for unbalanced loads. Each device consists of 21 separate control units, which are installed in Rittal enclosures. The outputs are galvanically connected to the power supply and can be adjusted independently of each other via three single-phase variable toroidal transformers with motor drive.

Each phase has its own controller, which controls the relevant control units in the laboratory power supplies independently of the other phases. The output voltage is set to the specified setpoint value separately for each phase via three external 0-10K potentiometers. This stabilises the set output voltage to an accuracy of +/- 1.0% regardless of mains fluctuations and load.

System for laboratory supply

Project Information

Installations for basic supply
Type: DAVRLU 32

Input 1:

3 x 400/230 V, 50 Hz, 3P, N, PE

Input 2

3 x 400/230 V, 60 Hz, 3P, N, PE

Output 1:

3 x 156/90 … 457/264 V, 50/60 Hz, 32 A

Output 2:

3 x 400/230 V, 50/60 Hz, 32 A

Output 3:

3 x 400/230 V, 60 Hz, 32A

Output 4:

3 x 400/230 V, 60 Hz, 16 A

Output 5:

230 V, 60 Hz, 16 A

Protection class:

1

Protection type:

IP21

Dimensions:

B 3000 x H 2000+100 x T 500 mm

Project Information

System for laboratory supply
Type: DSE 25-IS

Input=Output:

240/120 V, -15%/+15%, 60 Hz, L1, L2, N, PE switchable to 88 V, -15%/+15%, 60 Hz, L3, N, PE or 3 x 400/ 230 V, -15%/+ 15%, 50 Hz, L1, L2, L3, N, PE

Continuous current:

25 A, short-term 32 A

Setting accuracy:

+/- 0.7% of the final value for 240/120 V, 60 Hz +/- 1.5% of the final value for 88 V, 60 Hz +/- 0.7% of the final value for 3 x 400/230 V, 50 Hz

Setting speed:

15 V/s phase-phase for output 3 x 400/230 V
9 V/s phase-phase for output 240/120 V

Setpoint setting

via three potentiometers

Actual value acquisition:

via three measuring transformers with separate windings

Protection class:

1

Protection type:

IP21

Dimensions:

B 380 x H 600 x T 210 mm